User maintenance, Cleaning, Disinfecting – Shellab SRI20PF User Manual

Page 25

Advertising
background image

25 |

P a g e

USER MAINTENANCE

Warning

: Prior to any maintenance or service on this unit, disconnect the power cord from the power

supply.

Avertissement

: Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher le

cordon secteur de la source d'alimentation.

If a hazardous material/substance has spilled in the incubator

immediately initiate your site’s

Hazardous Material Spill Containment protocol. Contact your local Site Safety Officer and follow
instructions per the site policy and procedures.

Periodic cleaning and disinfection are required to prevent microbiological contamination.

C

LEANING

Note: The incubator chamber should be cleaned and disinfected prior to use.

Perform the following steps to clean incubator.

1. Remove all of the interior parts (shelves, racks, and any additional items), if assembled.

2. Clean the incubator with a mild soap and water solution, including all corners.

3. Rinse with distilled water and wipe dry with a soft cloth. Do not use deionized water.

4. Take special care when cleaning around the door alarm and temperature sensor heads,

and the chamber’s electrical outlet to prevent damage.

Warning

: Never clean the unit with alcohol or flammable cleaners.

Avertissement:

Ne jamais nettoyer l'appareil à l'alcool ou avec des nettoyants inflammables.

Note: Do not use spray cleaners or disinfectants that might leak through openings and cracks and

coat electrical components, or that contain solvents that will harm coatings. Do not use
chlorine-based bleaches or abrasives; they will damage the chamber liner.

D

ISINFECTING

Disinfect the incubator on a regular basis. Perform the below steps to disinfect the incubator:

1. Remove all of the interior parts (shelves, racks, and any additional items), if assembled.

Disinfect the Incubator, including all corners and the access port, using a suitable
disinfectant. Take special care when cleaning around sensing heads to prevent damage
and around the door gasket so as not to impair the positive seal.

2. Disinfect the incubator using commercially available disinfectants that are non-corrosive,

non-abrasive, and suitable for use on stainless steel, aluminum, painted steel surfaces.
Contact your local Site Safety Officer for detailed information for the proper disinfectants
suitable for your operation and laboratory protocol.