TEST FEATURE

After the press is turned on, press and hold the TEST button. While

depressing the TEST button, check the display. Make sure the bar graph

indicator on the display (a series of vertical lines that register to the 0 to

100% load meter scale – see example below) does not exceed 20%.

During the testing process, keep in mind that the ultrasonics are only

activated as long as the TEST button is depressed – once you release the

TEST button, ultrasonics is terminated.

A bar graph indicator reading of above 20%, signals that there may be a

problem with the stack. If this occurs check your assembly and re-test.

Frequency Display

The TEST key can also be used to display the running ultrasonic frequency.

Refer to page 19 for more information.

READY SCREENS

In addition to the start-up ready screen (shown on page 15), there are 2 other

“ready” screens. The press must be displaying one of the 3 ready screens in

order for welding to commence. Welding cannot be initiated from any other

display. The other 2 ready screens are as follows:

Alarm Screen

The Alarm screen displays information about the alarm(s) that have been

triggered by the system encountering parameters outside of the specified

tolerance limits. The screen will indicate whether alarms occurred in Time or

(on XPE models only) Energy. If there are no alarms, dashes will display. If

0045 Watts

NOTE: The TEST and Load
Meter check should always

be done for all cold start-

ups, and for any start-up

after the system has been

idle for 20 minutes or

more.

*Alarms*

Time: – –

Energy: – –