ThermoWorks P770 PRINTING DATALOGGER THERMOMETER User Manual

Page 11

Advertising
background image

Standard linearization / [oFF

Use the up and down keys [

ᮡᮢ

] to select [oFF.

Push [ENTER/MENUE] to confirm.

Calibration by code / oP1

Use the up and down keys [

ᮡᮢ

] to select oP1. Push

[ENTER/MENUE] to confirm the desired setting. On
the bottom of the display a small 1 appears, after
this number a four-digit number (Hex-Code /0..F) is
displayed. For changing the number use the up key
[

] . For stepping to the next number use the key

down [

] . If the requested number is complete

then push [ENTER/MENUE] to confirm. Now on the
bottom of the display appears a small 2 and after this
number a second four-digit number is displayed. For
changing the number please follow the manual as
before. Push [ESC] to be back in the measuring
mode.

1.4.5 Channel activation

(only 2-channel instruments) / [Chnl]

Chnl = channel

To activate or deactivate a measuring channel
push [ENTER/MENUE]. Use the up and down keys
[

ᮡᮢ

] to select Chnl. Push again [ENTER/MENUE].

On the left corner of the display appears a small 1,
which indicates the selected channel. For chang-
ing the channel use the up and down keys [

ᮡᮢ

].

Push [ENTER/MENUE] to confirm. Use the up and
down keys [

ᮡᮢ

] again to activate on or deactivate

off the requested measuring channel and push
[ENTER/MENUE] to confirm. Push [ESC] to be back
in the measuring mode.

Short-Cut: Press 2 seconds the HOLD/MAX/MIN-
Button for activate or deactivate channel 2.

Note: As a minimum one channel is active!

1.4.6 Enter Area for volume flow [ArEA]

To enter the area dimensions is only able with instru-
ments for measuring airflow.

CAL

oFF

oP1

oP2

11

Important!

Unit

MENUE

Prob

Lin2

CAL

PrEn Chnl

ArEA

Lo6