Δδ p – VACUUBRAND CVC 2II User Manual

Page 13

Advertising
background image

page 13 of 36

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 997835 / 15/12/2009

➨ Set hysteresis Δp by using keys Δp▲ or Δp▼:

☞ Set mode is activated by pressing key shortly: Set Δp ap-

pears.

☞ Short tips toggle in steps of 1 mbar.

☞ Continuous pressing activates ramp: Hysteresis Δp changes

with increasing speed.

Continuous pumping

Pumping down until the ultimate vacuum of the pump is
attained
(e. g. pump down to dry the system from volatile liq-
uids):
If the controller is in mode two-point control:
➨ Press key MODE to change to continuous pumping.

☞ Mode change is only possible if the process control is inac-

tive.

☞ The two-point control display disappears.

MODE

➨ Start: Press key START/STOP.

☞ Arrow down and valve and/or pump symbol are flashing.

➨ Stop: Press key START/STOP.

➨ Selecting the set point by using keys p▲ or p▼.

☞ Set mode is activated by pressing key shortly: Set p ap-

pears.

☞ Short tips toggle p in steps of 1 mbar.

☞ Continuous pressing activates ramp: set-point p changes

with increasing speed.

☞ Changing set point p leads at the same time to an automatic

preselection of hysteresis

Δp.

☞ Hysteresis Δp can be manually changed at any time similar

to p by using keys

Δp, see ”Manual setting of hysteresis

Δp”.

Two-point control

Manual setting of set point p (boiling point)

➨ Start continuous pumping.

☞ Switch over to continuous pumping by pressing key MODE

if controller is in two-point control mode.

☞ Mode change is only possible if process control is inactive.

Display two-point control disappears.

MODE

Semiautomatic determination of set point p (boiling point)

START

STOP

p

267

mbar

Δ

Δ

Δ

Δ

Δp

67

mbar

Manual setting of hysteresis

ΔΔΔΔΔp