How to change the basic setup

Starting the program:
➨ Press key MODE while switching on the controller.

☞ The basic setup as from last operation (factory-set: Stand-

ard) is displayed.

To change the basic setup:
➨ Press key p▲ or p▼ until the symbols of the desired basic

setup are displayed.

➨ Press key START/STOP to confirm the selected basic setup.

(mains switch)

MODE

Standard

An isolation valve is indispensable.

☞ Dispenses the cooling water according to the requirements

(cooling water valve 24 V= at the CVC 2

II

).

Management

Plus

An isolation valve and a VMS Module A are indispensa-

ble.

☞ Dispenses the cooling water according to the requirements.

☞ Process-orientated switching of the diaphragm pump in two-

point control operation and in continuous operation.

☞ High tolerance to condensates due to sufficient operation

time even in interval operation.

Management

A VMS Module A is indispensable.

☞ Process-orientated switching of the diaphragm pump in two-

point control operation and in continuous operation.

☞ Dispenses the cooling water according to the requirements.

☞ The pump is operated parallel (synchronous) to the isolation

valve.

☞ Energy consumption is reduced to a minimum.

☞ Operation is also possible without an isolation valve.

Attention: Take into account possible suck-back of gas (even
though at a very low rate) through the not-operating pump,
especially if the gas ballast valve is open. Make sure that no
potentially inflammable mixtures can be formed!

VACUU•LAN

A VMS Module A is indispensable.

☞ The vacuum pump is switched according to the preset pres-

sure and time values.

☞ The pump is switched off if no more pumping is required, the

pump is switched on again in case of pressure rise.

+