After setting the basic setup:
☞ The symbol of the cooling water valve is displayed and ”yes”

or ”no”.

➨ Press key p▲ or p▼ to activate (”yes”) or to inactivate (”no”)

the cooling water valve and press key START/STOP to con-
firm.

☞ The cooling water valve 24 V= at the controller is operated

parallel (synchronous) to the cooling water valve at the VMS
Module A.

☞ The controller is now ready for operation (process control inactive).

☞ The isolation valve is identified automatically.

☞ If it is missing in basic setup Standard or Management Plus, the process control cannot be started,

respectively is stopped.

☞ If it is missing in basic setup Management or VACUU•LAN, the valve symbol is not displayed.