Notes on operation

(1)

Connect the cable to the vacuum gauge head‘s serial in-

terface (MPT 100) or to its RS-485 socket (MPT 200) and

to one of the VACUU•BUS sockets (1) at the rear of the

vacuum gauge DCP 3000.

Attention: Do not cant when assembling or removing

plug connections! Comply with correct positioning of

the plug.

Only connect/disconnect the cable to/from the vacuum

gauge head when it is off circuit. Separate the DCP 3000

from mains.

Switch the DCP 3000 on.

When switching on the DCP 3000, a connected vacuum gauge head MPT 100 / MTP 200 is automati-

cally detected. The vacuum gauge head is displayed as active sensor in the status line of the vacuum

gauge as MPT 1. The pressure reading is indicated in exponential notation over the whole measuring

range.

Once all cables are connected, the vacuum gauge head MPT 100 / MPT 200 is ready to operate. Allow

the gauge head a stabilization time of approx. 5-10 minutes to meet its specifications. Even in case of

the vacuum gauge being switched off, the MPT 100 / MPT 200 stays ready for operation. To de-energize

the vacuum gauge head, unplug either its VACUU•BUS line or the power supply of the vacuum gauge.

The interior of the vacuum gauge head is highly sensitive! Do not insert fingers or

tools into the measuring chamber.

The measured pressure is dependent on the gas type (see ”Notes on operation”).

Max. permitted pressure at the pressure transducer: 1.5 bar (absolute).

Attention: Maximum pressure reading: 1*10

3

mbar. Pressure values above 1000

mbar are not displayed.

Risk of unnoticed overpressure. Risk of bursting!

NOTICE

Operating mode display via LED (only MPT 200)

Status of LED

Meaning

LED off

no or insufficient power supply

LED green on

valid measurement,

sensor component for low pressure range is active

LED green flashing

valid measurement,

sensor component for high pressure range is active

LED yellow on

unit function is O.K., but due to internal operations (e.g., during ad-

justment) no valid measuring values will be displayed temporarily

measuring range: overrange / underrange

LED red on

software or unit malfunction

LED green/yellow/red off (1s)

one-off: LED test after reset

LED red/green flashing

software update in progress