General information

To operate the vacuum gauge head MPT 100 / MPT 200, a vacuum

gauge DCP 3000 is required!

If using a different controller, order the instructions for use of the

MPT 100 / MPT 200 from VACUUBRAND.

Read this manual and the manual of the vacuum gauge DCP 3000 carefully be-

fore installing or operating the equipment. Observe the instructions contained

in these manuals.

Remove all packing material from the packing box. Remove the product from its

packing box and retain all packaging until the equipment is inspected and tested.

Inspect the equipment promptly and carefully.

If the equipment is damaged, notify the supplier and the carrier in writing within

three days. Retain all packing material for inspection. State the item number of the

product together with the order number and the supplier’s invoice number. Failure

to check and give notice of damage will void any and all warranty claims for those

deficiencies.

Store the equipment in dry and non-corrosive conditions if the equipment is not used

immediately.

Do not use any damaged equipment.

Intended use

Ensure that the individual components are only connected, combined and oper-

ated according to their design and as indicated in the instructions for use. Use

only original manufacturer’s spare parts and accessories. Otherwise the safety

and performance of the equipment, as well as the electromagnetic compatibility

of the equipment might be reduced.

The CE mark may be voided if not using original manufacturer’s spare parts.

Comply with all notes on correct vacuum and electrical connections, see section

”Use and operation”.

If installing the device into a cabinet or a housing check maximum temperature.

Install an external automatic ventilation system if necessary. Ensure that the

maximum permitted gas temperature at the sensor head (see ”Technical data”) is

not exceeded.

MPT 100:

The device is designed for

ambient and gas temperatures at the vacuum gauge

head connection of +10°C to +40°C at continuous operation.

MPT 200:

The device is designed for

ambient and gas temperatures at the vacuum gauge

head connection of +5°C to +60°C at continuous operation.