Guidance and manufacturer’s declaration – electromagnetic emissions

The ME equipment is intended for use in the electromagnetic environment specified below.

The customer or the user of the ME equipment should ensure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic environment – guidance

RF Emissions

CISPR 11

Group 1

The ME equipment uses RF energy only for its internal function. Therefore, its RF emissions are

very low and are not likely to cause any interference in nearby electronic equipment.

RF Emissions

CISPR 11

Class B

Complies

Harmonic emissions IEC

61000-3-2

Not Applicable

The ME equipment is

solely battery powered.

Voltage fluctuations /

flicker emissions

Not Applicable

Non-Life Support Equipment Separation Distance Calculation (3Vrms / 3V/m compliance)

Rated maximum output power of

transmitter (W)

Separation distance according to frequency of transmitter (m)

150 kHz to 80 MHz in ISM bands

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.17

1.17

2.33

10

3.69

3.69

7.38

100

11.67

11.67

23.33

Guidance and manufacturer’s declaration – electromagnetic immunity

The ME equipment is intended for use in the electromagnetic environment specified below.

The customer or the user of the ME equipment should ensure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

±6kV Contact

±8kV Air

Complies

Floors should be wood, concrete, or ceramic tile. If

floors are covered with synthetic material, the

relative humidity should be at least 30%.

Radiated RF

IEC 61000-4-3

3 V/m 80MHz to 2.5GHz

Complies

Field strengths outside the shielded location from

fixed RF transmitters, as determined by an

electromagnetic site survey, should be less than 3

V/m.

Interference may occur in the vicinity of equipment

marked with the following symbol:

Separation distance calculation provided above. If a

known transmitter is present the specific distance

can be calculated using the equations.

Conducted RF

IEC 61000-4-6

3Vrms 150kHz to 80MHz

Not Applicable

(no electrical cabling)

Electrical fast transient

IEC 61000-4-4

±2kV power line

±1kV I/O lines

Not Applicable

The ME equipment

is solely battery powered.

Surge

IEC 61000-4-5

±1kV differential

±2kV common

Not Applicable

Power frequency magnetic field

IEC 61000-4-8

3 A/m

Complies

Power frequency magnetic fields should be at levels

characteristic of a typical location in a typical

commercial or hospital environment.

Voltage dips, short interruptions

and voltage variations on power

supply input lines

IEC 61000-4-11

>95% dip 0.5 cycle

60% dip 5 cycles

70% dip 25 cycles

95% dip 5 sec.

Not Applicable

The ME equipment

is solely battery

powered.

d = [

3,5

] P

V

1

d = [

3,5

] P

E

1

d = [

7

] P

E

1