2 ground-referenced signal source, 3 signal connection, 1 voltage input mode – ADLINK USB-2401 User Manual

Page 38: Ground-referenced signal source, Signal connection, Voltage input mode, Figure 3-2, Ground-referenced source and differential input

Advertising
background image

28

Operation

3.2.2

Ground-Referenced Signal Source

Connected in some way to the existing ground system, to a
common ground point with respect to the USB DAQ, when the
computer is connected to the same power system. Non-isolated
output of instruments and devices connected to the existing
power systems are ground-referenced signal sources.

3.3

Signal Connection

Each analog input channel can be configured to different input
modes by the software API. Details of signal connection in
different input modes follow.

3.3.1

Voltage Input Mode

The properties of the signal to be measured must be considered.
The differential input mode provides two inputs that respond to
signal voltage difference between them. If the signal source is
ground-referenced, the differential mode can be used for the
common-mode noise rejection.

Connection of ground-referenced signal sources under differential
input mode is as shown.

Figure 3-2: Ground-Referenced Source and Differential Input

For floating signal sources, addition of a resistor at each channel
provides a bias return path. The resistor value should be about

AI-

AI+

Vcm

Common mode noise

& ground potential

Ground reference

signal source

GND

Amplifier

Amplifier

USB-2401