ADLINK ACL-8112 Series User Manual
Page 34
                                    Advertising
                                    
                                        
                                    
                                
                                 
26
• Signal Connections
AIHn
AILn
n = 0, ..., 8
Groun
Signal
Sourc
To A/D Converter
VG1
VG2
Vcm = VG1 - VG2
GND
+
-
Figure 3.5 Ground source and differential input
 
A differential mode must be used when the signal source is differential. A 
differential source means the ends of the signal are not grounded. To avoid 
dangerously high voltages between the local ground of the signal and the 
ground of the PC system, a shorted ground path must be connected. 
Figure 3.6 shows the connection for a differential source. 
 
AIHn
AILn
n = 0, ..., 8
Differentia
Signal
Sourc
To A/D Converter
VG1
VG2
Vcm = VG1 - VG2
GND
+
-
Figure 3.6 Differential source and differential input
                                    Advertising