Measuring Power or Loss (EPM-50)
8
4 Measuring Power or Loss
(EPM-50)
Nulling Electrical Offsets
Temperature and humidity variations affect the performance of electronic circuits 
and optical detectors. Nulling the electrical offsets eliminates these effects. Your unit 
has been designed not to require offset nulling under normal operation, but you 
should perform it whenever environmental conditions change significantly or when 
measuring very low power values.
Note:
Factory-defined values will be reinstated when you turn off the unit.
To perform an offset nulling:
Hold down 
and
a few seconds. The unit
displays NULL while nulling the offsets, then returns to 
normal mode.
Note:
Keypad is disabled during the operation.
I
MPORTANT
If light reaches the detector when nulling offsets, LIGH appears on 
the display and the nulling is not performed. You will need to press a 
key to return to the previous display.