EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual
Page 388

Referencia de instrucciones SCPI
382
FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:DISTance:FREQuency
Description
This command sets the frequency distance from
peak to center of noise region for noise
measurement of the WDM analysis default
channel.
At *RST, default channel noise measurement
distance is set to 100.0 GHz.
Syntax
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance<wsp><Distance[<wsp>HZ]>|MAXi
mum|MINimum|DEFault
Parameter(s)
Distance:
The program data syntax for <Distance> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is HZ. The
<Distance> special forms MINimum, MAXimum
and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.