Calibration[1..n]:zero[:auto – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual

Page 545

Advertising
background image

Referencia de instrucciones SCPI

OSA

539

Product-Specific Commands—Description

:CALibration[1..n]:ZERO[:AUTO]

Description

This command sets whether or not the

instrument should perform auto zero calibration

(nulling) at device-dependent intervals without

user intervention.

At *RST, auto zero calibration is set to on

(enabled).

Syntax

:CALibration[1..n]:ZERO[:AUTO]<wsp><Auto>

|ON|OFF|ONCE

Parameter(s)

Auto:
The program data syntax for <Auto> is defined

as a <Boolean Program Data>|<CHARACTER

PROGRAM DATA> element. The <Auto> special

forms ON and OFF are accepted on input for

increased readability. ON corresponds to 1 and

OFF corresponds to 0.

The parameter corresponds to the new state of

auto zero calibration.

0 or OFF: disable auto zero calibration.

1 or ON: enable auto zero calibration.

ONCE: launch a one time zero calibration. This

parameter has no effet on current ON/OFF state

of auto zero calibration.

Advertising
This manual is related to the following products: