Calibration[1..n]:zero[:auto – EXFO FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 User Manual
Page 546
Advertising

Referencia de instrucciones SCPI
540
FTB-5240S/S-P/BP
Product-Specific Commands—Description
Example(s)
STAT? Must return READY
CAL:ZERO:AUTO ONCE
STAT:OPER:BIT9:COND? Keep resending this
query as long as the operation is not complete
(returned value is not 0).
CAL:ZERO:AUTO? Returns 1 (auto zero still
enabled)
Notes
Zero calibration operation takes up to 5 seconds
to complete.
Auto zero calibration cannot be disabled: OFF (0)
value is valid for queries only.
See Also
:CALibration:ZERO:AUTO?
:STATus?
:STATus:OPERation:BIT<8|9>:CONDition?
:CALibration[1..n]:ZERO[:AUTO]