EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 116

Advertising
background image

Testing Fibers in Template Mode

102

OTDR

Acquiring Traces in Template Mode

3d. If you want to automatically apply the fiber span defined in the

template reference trace to all acquired traces, select the
Apply reference fiber span to current traces during
post-processing
check box.

If you clear the box, the analysis will be performed on the
common portion of the areas delimited by the span start and end
of the reference trace and the span start and end of the main
trace.

3e. Press Apply to confirm, then OK to return to the main window.

Once Template mode is selected, the reference trace is displayed
in red on the graph.

3f. If you selected Reference Update at step 4b, update your

reference trace as follows:

3g. Press Start.

If the first connector check feature is enabled, a message will
appear if there is a problem with the injection level (see Enabling
or Disabling the First Connector Check
on page 54).

All traces will automatically be acquired and analyzed, and the
events will be identified.

Note: In offline operation, instead of pressing Start to acquire traces, you simply

recall traces stored on the hard disk of the FTB-500.