SPECIFICATIONS

Range

0.0 to 50.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.5 mg/L ±5% of reading at 25 °C

Typical EMC

±0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, D1426-92,

Nessler method. The reaction between ammonia and reagents causes a yellow tint in

the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93733A-0

Nessler Reagent 4 drops

(in fresh and seawater)

HI 93733B-0

Ammonia Reagent 9 mL

(in fresh and seawater)

REAGENT SETS

HI 93733-01 Reagents for 100 tests
HI 93733-03 Reagents for 300 tests
For other accessories see page 34.

MEASUREMENT PROCEDURE

• Select the

Ammonia HR method using the procedure described in the

Method Selection section (see page 11).

• Fill a cuvette with 1 mL of unreacted sample, by means of the

syringe.

• Add 9 mL of HI 93733B-0 Ammonia Reagent, by means of the 3 mL

plastic pipette. Place the cap and swirl the solution to mix.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

AMMONIA HIGH RANGE

Ammonia HR

1 mL

9 mL