Nitrite HR

NITRITE HIGH RANGE

SPECIFICATIONS

Range

0 to 150 mg/L

Resolution

1 mg/L

Accuracy

±4 mg/L ±4% of reading at 25 °C

Typical EMC

±1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the Ferrous Sulfate method. The reaction between nitrite and the
reagent causes a greenish-brown tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93708-0

Powder reagent

1 packet

REAGENT SETS

HI 93708-01 Reagents for 100 tests
HI 93708-03 Reagents for 300 tests
For other accessories see page 34.

MEASUREMENT PROCEDURE

• Select the

Nitrite HR method using the procedure described in the Method Selection section (see

page 11).

• Fill the cuvette up to the mark with 10 mL of unreacted

sample and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette.