Hanna Instruments HI 83218-01 User Manual

Page 28

Advertising
background image

28

10 mL

SPECIFICATIONS

Range

0.00 to 1.15 mg/L

Resolution

0.01 mg/L

Accuracy

±0.06 mg/L ±4% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

EPA Diazotization method 354.1. The reaction between nitrite and

the reagent causes a pink tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93707-0

Powder reagent

1 packet

REAGENT SETS

HI 93707-01 Reagents for 100 tests
HI 93707-03 Reagents for 300 tests
For other accessories see page 34.

MEASUREMENT PROCEDURE

• Select the

Nitrite LR method using the procedure described in the Method Selection section (see

page 11).

• Fill the cuvette up to the mark with 10 mL of unreacted sample

(up to the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette.

Nitrite LR

NITRITE LOW RANGE