• Remove the cuvette.

• Add 4 drops of HI 93715A-0 First Reagent (6 drops for

seawater analysis). Replace the cap and mix the solution.

• Add 4 drops of HI 93715B-0 Second Reagent (10 drops

for seawater analysis). Replace the cap and mix the solution.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the

measurement or, alternatively, wait for 3 minutes and 30 seconds
and press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L of ammonia nitrogen (NH

3

-N).

• Press  or  to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of ammonia (NH

3

) and ammonium (NH

4

+

).

• Press  or  to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic chloramines, sulfide, various

aliphatic and aromatic amines.

Ammonia MR