General tips for an accurate measurement – Hanna Instruments HI 96728 User Manual

Page 11

Advertising
background image

11

The instructions listed below should be carefully followed during testing to ensure best accuracy.

• Color or suspended matter in large amounts may cause

interference, therefore these should be removed by
treatment with active carbon and by prior filtration.

• For a correct filling of the cuvette: the liquid in the

cuvette forms a concavity on the top; the bottom of this
concavity must be at the same level of the 10 mL mark.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;

(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

• It is important that the sample does not contain any debris. This would corrupt the reading.

• Each time the cuvette is used, the cap must be tightened to the same degree.

• Whenever the cuvette is placed into the measurement cell, it

must be dry outside, and completely free of fingerprints, oil
or dirt. Wipe it thoroughly with HI 731318 or a lint-free
cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain

accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.
• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvette for zeroing and measurement.

Advertising