Hanna Instruments HI 96728 User Manual

Page 13

Advertising
background image

13

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will show

“-0.0-”

. The meter is now zeroed and

ready for measurement. Remove the cuvette.

• Add one packet of the HI 93728-0

reagent.

• Replace the cap and immediately shake

vigorously for exactly 10 seconds by
moving the cuvette up and down. Continue
to mix by inverting the cuvette gently
and slowly for 50 seconds, while taking
care not to induce air bubbles. A deposit
could remain but it dose not affect the
measurement. Time and way of shaking
could sensitively affect the measurement.

• Replace the cuvette into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER for three

seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.

Advertising