Errors and warnings – Hanna Instruments HI 96728 User Manual

Page 9

Advertising
background image

9

ERRORS AND WARNINGS

The instrument shows clear messages when erroneous condition appears. Messages are also
displayed when the obtained values are outside expected range. The beeper is playing a beep on
errors.

a) on zero reading

Light High:

There is too much light to perform a measurement.

Please check the preparation of the zero cuvette.

Light Low:

There is not enough light to perform a measurement.

Please check the preparation of the zero cuvette.

No Light:

The instrument cannot adjust the light level. Please

check that the sample does not contain any debris.

Inverted cuvettes:

The sample and the zero cuvette are inverted.

Zero:

A zero reading was not taken. Follow the instructions of

the measurement procedure for zeroing the meter.

b) on sample reading

Under Range:

A blinking “0.0” indicates that the sample

absorbs less light than the zero reference. Check the procedure
and make sure you use the same cuvette for reference (zero) and
measurement.