Specifications precision and accuracy – Hanna Instruments HI 96728 User Manual

Page 5

Advertising
background image

5

SPECIFICATIONS

PRECISION AND ACCURACY

Range

0.0 to 30.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.5 mg/L ±10% of reading@ 25°C

Typical EMC Deviation

±0.1 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525 nm

Method

Adaptation of the cadmium reduction method. The reaction between nitrate-
nitrogen and the reagent causes an amber tint in the sample.

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode

;

after 1 hour of non-use in

calibration mode

;

with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7 ”)

Weight

360 g (12.7 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93728-0

Powder Reagent

1 packet

Precision

is how closely repeated measurements

agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy

is defined as the nearness of a test

result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of
10.0 mg/L nitrate and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 0.3 mg/L.

Advertising