English, Manual (p. 2), Blood pressure monitor – Konig Electronic Blood pressure monitor wrist User Manual

Page 2: Specifications, Notice before measurement, Battery loading

Advertising
background image

2

ENGLISH

Blood Pressure

Level Classification

Indicator

Memory Button/

Clock Adjusting

LCD

Power Switch/

Clock Setting

SPECIFICATIONS

1. Model: HC-BLDPRESS11
2. Classification: Internally powered, type B
3. Machine size: 81mm×71mm×30mm
4. Cuff size: 300mm×70mm
5. Weight: about 120g (exclude batteries)
6. Measuring method: oscillometric method, automatic inflation and measurement
7. Memory volume: 60 times with time and date stamp
8. Power source: AAA×2 batteries, Alkaline type
9. Cuff pressure range: 0-300mmHg
10. Pressure accuracy: ±3mmHg
11. Pulse rate range: 40-180/min
12. Environmental temperature for operation: 5°C~40°C
13. Environmental humidity for operation: <90%
14. Environmental temperature for storage: -20°C~55°C
15. Environmental humidity for storage: <95%
16. Environmental pressure: atmosphere pressure
17. Battery life: Approx.1 month with normal usage

NOTICE BEFORE MEASUREMENT

(1) Stay quiet, calm and rest for 5 minutes before blood pressure measurement.
(2) The cuff should be placed at the same level as your heart.
(3) During measurement, neither speak nor move your body and arm.
(4) Measuring on same wrist for each measurement.
(5) Please always relax a minimum of 3 to 5 minutes between measurements to allow the blood circulation in your arm to

recover.

(6) Remove the batteries if the monitor will not be used for a month or more to avoid damage or battery leakage.
(7) This blood pressure monitor is designed for use with adults and should never be used for infants or young children.

Consult your physician or other health care professionals before using on older children.

(8) The cuff size is suitable for adult’s wrist and this model is intended to be used by adults only.
(9) Persons with arrhythmia are not acceptable for this blood pressure monitor.
(10) Please wear the cuff on the left wrist.
(11) Please avoid strong magnetism interference, such as mobile telephone, microwave oven, etc.

1. BATTERY LOADING

a. Open battery cover at the back of the monitor.
b. Load two “AAA” size batteries. Please pay attention to polarity.

Advertising