9. TROUBLESHOOTING

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD Display

shows battery

symbol

Low Battery

Change all batteries

LCD Display

shows “EE”

Arm or blood pressure monitor was moved

during testing

Re-test taking care to not move your arm or the blood

pressure monitor

The cuff does not inflate properly or pressure

falls quickly during testing

Ensure that the rubber tube is fully inserted into the

blood pressure monitor

Irregular heartbeat (arrhythmia)

People with light arrhythmia may re-test.
It is inappropriate for people with serious arrhythmia

to use this blood pressure monitor.

The blood pressure monitor did not inflate

sufficiently.
The cuff was not properly wrapped or the

rubber tube was bent or pressed.

Review the cuff wrapping and testing sections of the

instructions and re-test.

LCD Display

shows abnormal

result

The cuff position was not correct or it was not

properly tightened

Apply the cuff correctly and try again

Body posture was not correct during testing

Review the body posture and testing sections of the

instructions and re-test.

Speaking, arm or body movement, angry,

excited or nervous during testing

Re-test when calm and without speaking or moving

during the test

No response when

you press button

or load battery

Incorrect operation, or strong electromagnetic

interference

Take out batteries for five minutes, and then reinstall

all batteries.

MAINTENANCE

1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and polarization. Do not immerse the monitor in water as this will result in damage to the

monitor.

3. If this monitor is stored near freezing, allow it to acclimate to room temperature before use.
4. Do not attempt to disassemble this monitor.
5. If you do not use the monitor for a long time, please remove the batteries.

When the cuff is dirty, detach it from the equipment, wash the cuff by hand with proper detergent and rinse in

cold water. Dry in air. Never iron it.

EXPLANATION OF SYMBOLS ON UNIT

Symbol for “Caution Consult Accompanying Documents”

Symbol for “Type B Applied Parts”

CE is a kind of marking of EU safe attestation. This product matches MDD93/42/EEC

Safety precautions:

To reduce risk of electric shock, this product should ONLY be opened by an authorized

technician when service is required. Disconnect the product from mains and other equipment if a

problem should occur. Do not expose the product to water or moisture.

Maintenance:

Clean only with a dry cloth. Do not use cleaning solvents or abrasives.

Warranty:

No guarantee or liability can be accepted for any changes and modifications of the product or damage caused due to

incorrect use of this product.

RISK OF ELECTRIC SHOCK

DO NOT OPEN

CAUTION