9 meter mode, Front panel interface 23 – Lab.gruppen LM 44 User Manual
Page 29

Front Panel Interface
23
Lake LM Series Operation Manual Rev 1.2.8
6.8.2 Clip Indication
The dedicated 5-segment metering LEDs (Figure 6-4 ) display Module input and output clip or pre-clip
conditions. Additionally, the faults and warnings LED described in section 6.7.2 provides clip warnings for
input mutes, or GPIO protective mutes, along with confirmation text on the LCD screen.
Clipping is monitored at the following positions in the signal chain:
▸
Analog Inputs: If the input signal exceeds either +12 dBu or +26 dBu (according to analog sensitivity
setting), a clip indication will be given. This does not apply if digital inputs are selected.
▸
Module Inputs: If the signal level at this point exceeds +25 dBu, a clip indication will be given.
▸
Module Outputs: If the signal level at this point exceeds +21 dBu, a clip indication will be given.
6.9 Meter Mode
6.9.1 Home View (Contour Mode)
The default view when powering on an LM Series device device configured in Contour Mode is the Meter
Mode > Home View as shown in Figure 6-7.
Figure 6-7: Meter Mode > Home View
Home View (Contour Mode) provides a summary of Module I/O gain level and limiter gain reduction, along
with frame, module and channel labeling information.
6.9.2 Home View (Mesa Mode)
The default view when powering on an LM Series device configured in Mesa Mode is Meter Mode > Home
View as shown in Figure 6-8.