Labconco XPert Weigh Boxes 5220331 User Manual

Page 46

Advertising
background image

Product Service 1-800-522-7658

42





C

C

h

h

a

a

p

p

t

t

e

e

r

r

8

8

:

:

T

T

r

r

o

o

u

u

b

b

l

l

e

e

s

s

h

h

o

o

o

o

t

t

i

i

n

n

g

g

Refer to the following table if the XPert filtered glove box fails to operate
properly. If the suggested corrective actions do not solve your problem,
contact Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Blower won’t operate.

Unit not plugged into
outlet.

Plug the glove box into appropriate electrical
service.

Circuit

breaker(s)

or

Ground Fault Interrupter.

Reset circuit breaker.

Blower

wiring

is

disconnected.

Inspect blower wiring per wiring diagram on
unit.

Blower

switch

is

defective.

Replace switch if no voltage at blower.

Motorized impeller or
blower is defective.

Replace motorized impeller or blower. See
Chapter 6.

Unit not plugged into
outlet.

Plug unit into appropriate electrical service.

Blower and lights won’t
operate.

Circuit breaker(s) tripped. Reset or replace circuit breaker.

Lights do not work.

Lamp not installed
properly.

Inspect lamp installation.

Lamp

wiring

disconnected.

Inspect lamp wiring.

Defective lamp.

Replace lamp.

Light switch is defective.

Replace light switch.

Defective

electronic

ballast.

Replace ballast.


Advertising