Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

29

Information Center

The Information Center is an LCD display located on the right side wall at
eye level. The center’s display is divided into three separate areas, as shown
in Figure 5-2, below. The display will turn itself off one minute after the
blower is turned off or the sash is closed.

Status Area

Data Area

Icon Area

Status Area

The Status Area displays the system’s condition during operation, as well as
the inflow velocity. This line can display any of the following messages:

Blower Off

The cabinet blower is shut off.


Please Wait

For three minutes after the cabinet blower is turned on, the display will
say this, to allow the cabinet to reach proper operating conditions.


OK

The motor is operating properly, and the sash is not open too far.


Sash is too high

The sash is open too far for safe operation.


Airflow Alert

The airflow patterns in the cabinet have changed, resulting in a sudden
change in the motor speed. This is most likely due to a blockage of the
prefilter.

Figure 5-2

Advertising