Labconco Biological Safety Cabinets 34610 Series User Manual

Page 7

Advertising
background image

Product Service 1-800-522-7658

2









C

C

h

h

a

a

p

p

t

t

e

e

r

r

2

2

:

:

P

P

r

r

e

e

r

r

e

e

q

q

u

u

i

i

s

s

i

i

t

t

e

e

s

s

Before you install the Biosafety Cabinet, you need to prepare the site for
installation. Examine the location where you intend to install the cabinet.
You must be certain that the area is level and of solid construction. In
addition, a dedicated source of electrical power must be located near the
installation site.

Carefully read this chapter to learn:

• Location requirements.

• Electrical power requirements.

• Exhaust requirements.

• Service utility requirements.

• Space requirements.

• User-supplied microscope requirements (if needed)

• User-supplied circulator requirements (if needed)

Refer to Appendix C: Specifications, for complete biosafety cabinet electrical
and environmental conditions, specifications and requirements.

Space Requirements

The dimensions for the Cell Logic B2 Biosafety Cabinets are shown in Appendix
B: Dimensions.

Clearance

A minimum clearance of at least 6 inches (150 mm) is suggested on the top
and both sides of the cabinet for service.

Advertising
This manual is related to the following products: