The logic touchpad – Labconco Biological Safety Cabinets 34610 Series User Manual

Page 38

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

33

The Logic Touchpad

The touchpad of the Logic Series Cabinet is shown in Figure 5-4. Take a
moment to get familiar with the buttons, their location and function. Also
familiarize yourself with the display located on the right side wall. The
display will report system functions, such as filter capacity, timer displays,
alarm or error messages, as well as icons that illuminate when cabinet
functions such as UV light and blower are operational


Blower Button – Starts or stops the cabinet blower. When
the blower is in Smart-Start

TM

mode, closing the sash turns

the blower off. When the sash is raised, the blower restarts
automatically. Pressing this button overrides Smart-Start

TM

operation.

Light Button – Turns the fluorescent lamps on or off.
Closing the sash automatically turns the lights off. When
the lights are in Smart-Start

TM

mode, raising the sash turns

the lights on automatically.

UV Light Button – Turns the UV lamp on or off. When
the UV lamp is in Smart-Start

TM

mode, closing the sash

turns the light on. When the sash is raised, the light turns
off automatically.

Timer Button – Allows you to select either a repeating
interval timer, or an elapsed timer (stopwatch).

Mute/OK Button – Mutes all audible alarms for
approximately 5 minutes, unless it is a system error alarm.
When in the Menu mode, this button is used to select an
option.

Menu Button – This button toggles the display between
the display and menu modes. When in the menu mode,
pressing this button returns you to the previous menu level.

Select Buttons – Allow you to choose different options in
the menu mode.

Figure 5-4

Advertising
This manual is related to the following products: