Chapter 7: Troubleshooting
Product Service 1-800-522-7658
50
PROBLEM
CAUSE
CORRECTIVE ACTION
Fluorescent light not 
working 
Sash is closed
Open sash – Fluorescent lights will not 
work with the sash closed. 
Lamp(s) are defective Replace defective lamp(s)
Lamp wiring is 
disconnected 
Inspect lamp wiring.
Defective
lamp
ballasts
Replace lamp ballasts.
Keypad
disconnected
or defective
Run keypad diagnostics and check 
connections. 
Fluorescent light is 
dim or flickering 
Lamp(s) are defective Replace defective lamp(s)
Lamp wiring is 
disconnected 
Inspect lamp wiring.
Defective lamp 
ballast 
Replace lamp ballast.
UV light not 
working 
Sash is open
Close sash – UV light will not work 
with the sash open. 
Lamp is defective
Replace defective lamp.
Lamp wiring is 
disconnected 
Inspect lamp wiring.
Defective
lamp
ballast
Replace lamp ballast.
Keypad
disconnected
or defective
Run keypad diagnostics and check 
connections. 
UV light is dim or 
flickering 
Lamp is defective or 
is at end of operating 
lifetime. 
Replace defective or worn out lamp.
Lamp wiring is 
disconnected 
Inspect lamp wiring.
Defective lamp 
ballast 
Replace lamp ballast.