Chapter 2: Prerequisites
Product Service 1-800-522-7658
3
Location Requirements
Note: The Biosafety Cabinet should be located away from traffic patterns, 
doors, fans, ventilation registers, fume hoods and any other air-handling 
devices that could disrupt its airflow patterns. All windows in the room 
should remain closed. Figure 2-1 shows the preferred location for the 
Biosafety Cabinet. 
 
 
Preferred location 
Air register blocked or redirected
to prevent cabinet disruption
Alternate location
 
 
 
 
 
 
 
 
 
 
Figure 2-1