Using the system – Labconco CAptureTMPortable Fuming System 3420024 User Manual

Page 16

Advertising
background image

Chapter 4: Using the System

Product Service 1-800-522-7658

12

Using the System

Set Up

After the Portable Fuming System has been removed from its carton and
positioned in a suitable location, you must:

1. Release the two clamps

securing the transport cover to
the clear chamber, as shown in
Figure 4-3. Set the transport
cover aside. Lift the heater/filter
module out of the chamber.
Remove the accessories and the
foam storage insert from the
bottom of the clear chamber.

2. Put the leveling foot in the down

position by pulling the pin out and
sliding the foot down until the pin sets,
as shown in Figure 4-4.

1. Before using the system, protect the

inside surface of the clear chamber with
Rain-X® or Rain Block

TM

glass water

repellent. Labconco has tested these
two treatments, and found they protect
the clear chamber from CA fumes for
up to 20 uses, if the treatment is not
disturbed. Always follow the treatment
manufacturer’s instructions when
applying the treatment.

Figure 4-3

Figure 4-4

Advertising