The common service operations necessary to maintain the CApture Portable
Fuming System’s peak performance are listed below.

Routine Maintenance Schedule

Weekly

• Using an appropriate soap or detergent solution, or LabSolutions

Glass & Surface Wipes, Labconco part # 1570000; clean the stainless
steel face plate.

• If the inside surface of the clear chamber is cleaned, reapply either

Rain-X® or Rain Block treatment

TM

.

Monthly (or more often as required)

• Using a damp cloth, or LabSolutions Glass & Surface Wipes,

Labconco part # 1570000; clean the exterior surfaces of the system, to
remove any accumulated dust.

• All weekly activities.

After every 20 development cycles

• Carefully clean the clear chamber with mild detergent or glass

cleaner. When thoroughly clean and dry, treat the inside of the
chamber with Rain-X® or Rain Block treatment

TM

.

After every 100 development cycles

• Replace the filter cartridge.