Chapter 2: Prerequisites

Product Service 1-800-522-7658

3

Exhaust Requirements

If you intend to connect the system to the optional Exhaust Connection Kit,
first examine the location to ensure that the exhaust duct can be directed
either out of the immediate area (through a partially opened window), or into
a chemical fume hood. For further information about the CApture’s exhaust
system requirements, please refer to Appendix D: Accessories.

Note: For best results, do not add any additional exhaust duct to that already
supplied in the exhaust connection kit. Additional lengths of duct will reduce
the exhaust flow out of the chamber, possibly resulting in incomplete venting
of the CA fumes.

Electrical Requirements

The system models have the following electrical requirements:

Model #

Requirements

3420000

115 VAC, 60 Hz, 1 Amp

34200-20, -22, -24

230 VAC, 50/60 Hz, 0.5 Amp

Note: An outlet with a circuit breaker rated at 10 amps for 115 volt
models (5 amps for 230 volt models) should be located as close as possible
to the system. Power cords supplied with the print system are 3 meters
(10 feet) long.

Note: The system can be operated on an inverter, converting 12 Volts DC
to the appropriate AC voltage. The inverter should have an output
rating of at least 100 Watts, and protected against current overload for
proper operation.