Chapter 3: Getting Started

Product Service 1-800-522-7658

10

Electrical Connection

Plug the power cord into the receptacle on the back of the CentriVap Concentrator
and plug the other end into a suitable power receptacle. Plug the power cord into
the receptacle on the back of the CentriVap Cold Trap and plug the other end into
a suitable power receptacle. Plug the power cord from the vacuum pump into the
receptacle on the back of the Concentrator. If the vacuum pump has an off/on
switch, turn the switch ON. The vacuum pump will be controlled by the
Concentrator.

Rotor Installation

Place the rotor onto the shaft of the Concentrator. Rotate the rotor slightly to
engage the drive pin in the shaft with the slots in the rotor hub. The top of the
shaft should be in line with the top of the rotor hub.

IMPORTANT NOTE: Do not use a rotor if it shows any signs of damage.
Failure of a spinning rotor could damage the CentriVap or cause samples to
be lost.

Chemical Resistance of CentriVap Components

Your CentriVap Centrifugal Concentrator and Cold Trap are designed to be
chemical resistant to most compounds that are commonly used in concentration
processes. However, by necessity, the CentriVap is comprised of a number of
different materials, some of which may be attacked and degraded by corrosive
chemicals. The degree of degradation is obviously dependent on the concentration
and duration of exposure. Some major components of the CentriVap that are
susceptible to degradation are as follows: