Chapter 4: Using Your CentriVap Refrigerated Concentrator

Product Service 1-800-522-7658

17

Change “Run Time” set points:

To change the time set points, press the set point “SELECT” button until arrows
point to the “Time at 1” or “Time at 2” set point which can be changed from 1 to
999 minutes by pressing the increase or decrease buttons. If it is desired to have
the CentriVap run continuously without alarming at the end of a time period,
press the increase button until the “Time” display says “ON.” The last entered set
point is stored in memory.

If during a run the “STOP” button is pressed, the timers remember the time at
which stop occurred. If “RUN” is then pressed, the timers continue to count
down from the time at which they were stopped. If running program 1, program 2
or program 3, pressing “PROG 1,” “PROG 2” or “PROG 3” resets the timers to
the original set point time and the CentriVap starts a new run.
To reset the timers to the original set point time when running programs other
than program 1, program 2 or program 3, press the start button and hold it for five
seconds. The display will indicate that the timer is reset.

Operating the CentriVap

CAUTION: To avoid injury; Do not operate the CentriVap if the lid is
scratched, nicked or shows signs of damage. A damaged lid could fail under
vacuum.


While the CentriVap Concentrator is operating, do not lean on the lid, do not
stand near it longer than necessary and do not place hazardous materials
within 12 inches.
1. Press the Cold Trap “ON” switch. The top amber indicator will illuminate.

Additional indicators will illuminate, as the Cold Trap gets colder. A green
indicator will illuminate when the Cold Trap reaches operating temperature.

2. Press the CentriVap “ON” switch.
3. Select a program or set the set point parameters.
4. Preheat/precool the chamber if desired.
5. Place samples in vials. Normally the vials should be no more than half full.

Place vials in a rotor.