Chapter 7: troubleshooting – Labconco CentriVap Refrigerated Centrifugal Concentrators 7310000 Series User Manual

Page 31

Advertising
background image

Chapter 7: Troubleshooting

Product Service 1-800-522-7658

27


Other corrective actions for potential problems are as follows:

PROBLEM

CAUSE

CORRECTIVE ACTION

Unit will not operate

Unit not connected to electrical
power

Connect unit to proper electrical receptacle.

Circuit breaker blown

Correct electrical problem and reset circuit
breaker by pressing button.

Lid open

Close lid.

Excessive vibration

Sample tubes not located
symmetrically in rotor

Reposition sample tubes.

Sample odor in lab

Vent hose exhausting into lab
area

Redirect hose to fume hood.

Evaporation rate is reduced

Heater inoperable

Contact Labconco.

Vacuum pump failure

Check pump.

Obstruction in hose

Remove obstruction or replace hose.

Lack of adequate vacuum

See below.

No vacuum/poor vacuum

Pump not on

Turn on pump.

Control valve open

Check control valve.

Leaks in lines or connectors or
gasket

Locate and repair.

Foreign material on lid gasket

Clean gasket and lid.

Pump is not functioning properly Check pump by locating vacuum gauge closer to

pump and close off rest of system. Check pump
oil for cloudiness or particles and change.

If pump is faulty, seek authorized service or
replace pump.

Cold Trap or Concentrator
gasket is not sealing properly

Check gasket for cleanliness. Adjust gasket as
needed.

Cold Trap lid not seated

Hold lid down until vacuum is initiated.

Ice formed on Cold Trap lid
sealing surface

Defrost and wipe dry.

Secondary Trap Canister not
fitted properly

Tighten all connections to and from the
Secondary Trap Canister.

Secondary Trap Insert is spent

Replace with new insert.

New Secondary Trap has
moisture in it

Run vacuum pump for 24 hours to remove
moisture.