Location requirements, Vacuum pump requirements, Chamber or manifold requirements – Labconco FreeZone 6 Liter Benchtop Freeze Dry Systems 77520 Series User Manual

Page 12

Advertising
background image

8

Chapter 2: Prerequisites

Product Service: Domestic 1-800-522-7658, International 816-333-8811

Location Requirements

The Freeze Dryer should be located in an area that provides an
unobstructed flow of air around the cabinet. This air cools the
refrigeration system. The refrigeration system draws air in through
the rear and exhausts it through the sides. A minimum of 3" must
be allowed between the rear and both sides of the Freeze Dryer and
adjacent wall surfaces. Restriction of airflow during operation
could adversely affect performance.

Refer to Appendix C: Freeze Dryer Specifications for dimensional
drawings of the Freeze Dryer. The space for benchtop models
must be large enough to accommodate the vacuum pump, which
will be used with the Freeze Dryer.

Vacuum Pump Requirements

A vacuum pump must be provided by the user. A vacuum pump
with a displacement of 144 liters per minute and 0.002 mBar
ultimate pressure is adequate for most samples. The inlet fitting on
the vacuum pump must be suitable for 3/4" ID vacuum hose,
which is provided. It is recommended that the vacuum pump is
equipped with an exhaust filter to minimize oil mist exhausting
from the vacuum pump. The operating vacuum level may be set
on the Freeze Dryer. The higher the pressure is set, the more likely
it is that oil mist will be exhausted.

Vacuum pumps used with 115V models should be equipped with
an 115V, 15 Amp NEMA 5-15P plug. Vacuum pumps used with
230V models should be equipped with a reverse IEC plug. This
plug is included with 230V models. This will allow the vacuum
pump to be plugged into the receptacle on the back panel of the
freeze dryer. See Appendix D: Freeze Dryer Accessories for
vacuum pumps available from Labconco.

Chamber or Manifold
Requirements

A freeze drying chamber is not included and must be purchased
separately. These allow samples to be attached to the Freeze
Dryer. See Appendix D: Freeze Dryer Accessories for available
chambers and manifolds.

Advertising