Chapter 7: Troubleshooting

Product Service: Domestic 1-800-522-7658, International 816-333-8811
























III. System Components and Collection
Chamber Isolation

1. Remove the drying chamber or manifold.
2. Leave the gasket on and turn a large freeze dry flask upside

down to cover the connection port.

3. Start the freeze dryer and observe the vacuum indication.

If the vacuum indication is good, then the problem is in the
drying chamber or manifold and you should return to
Section II, Gaskets, Tubing, Connections, Sample Valves.










Thrust Washer

Valve Body

Knob & Stem

Assembly

Trim Ring

Manifold Assembly

Knob & Stem

Assembly

Bulkhead Nut

Gasket
(No grease
required)

Bulkhead

Fitting

Valve Body

Valve Assembly (Manifold)

Valve Assembly (Stainless Steel Chamber)