Labconco FreeZone 6 Liter Benchtop Freeze Dry Systems 77520 Series User Manual

Page 56

Advertising
background image

52

Appendix C: Freeze Dryer Specifications

Product Service: Domestic 1-800-522-7658, International 816-333-8811



Environmental Conditions

• Indoor use only.

• Maximum altitude: 6562 feet (2000 meters).

• Ambient temperature range: 41° to 104°F (5° to 40°C).

• Maximum relative humidity: 80% for temperatures up to

88°F (31°C), decreasing linearly to 50% relative humidity
at 104°F (40°C).

• Main supply voltage fluctuations not to exceed ±10% of the

nominal voltage.

• Transient over voltages according to Installation Categories

II (Over voltage Categories per IEC 1010). Temporary
voltage spikes on the AC input line that may be as high as
1500V for 115V models and 2500V for 230V models are
allowed.

• Used in an environment of Pollution degrees 2 (i.e., where

normally only non-conductive atmospheres are present).
Occasionally, however, a temporary conductivity caused by
condensation must be expected, in accordance with IEC
664.


Advertising