Appendix C: Freeze Dryer Specifications

Product Service 816-333-8811 or 1-800-522-7658

48

Catalog #

Voltage

Frequency

Phase

Amperage

Nominal

Operating

Range

Nominal

w/o

Vacuum

Pump

Max. with

Vacuum

Pump

7382020

115V

103-127

60

1

11.5

16

7382021

115V

103-127

60

1

11.5

16

7382030

230V

198-254

50

1

6.3

10

7382031

230V

198-254

50

1

6.3

10

7382032

230V

198-254

50

1

6.3

10

7382033

230V

198-254

50

1

6.3

10

7382034

230V

198-254

50

1

6.3

10

7382035

230V

198-254

50

1

6.3

10

7382040

230V

187-254

60

1

6.3

10

7382041

230V

187-254

60

1

6.3

10


Environmental Conditions

Indoor use only.

Maximum altitude: 6562 feet (2000 meters).

Ambient temperature range: 41° to 104°F (5° to 40°C).

Maximum relative humidity: 80% for temperatures up to 88°F

(31°C), decreasing linearly to 50% relative humidity at 104°F (40°C).

Main supply voltage fluctuations not to exceed ±10% of the nominal

voltage.

Transient overvoltages according to Installation Categories II

(Overvoltage Categories per IEC 1010). Temporary voltage spikes on
the AC input line that may be as high as 1500V for 115V models and
2500V for 230V models are allowed.

Used in an environment of Pollution degrees 2 (i.e., where normally

only non-conductive atmospheres are present). Occasionally,
however, a temporary conductivity caused by condensation must be
expected, in accordance with IEC 664.

Heat rejected from freeze dyer excluding the vacuum pump: Less

than 4320 BTU/HR.