Note: This Quick Guide to Installation may not provide all the information for

your installation. Refer to the User’s Manual for details and precautions. Do not

discard packaging until you have identified all parts that were shipped and the

freeze dryer operates properly.

P

P

r

r

e

e

p

p

a

a

r

r

a

a

t

t

i

i

o

o

n

n

s

s

Step 1. Proper Electrical Supply

• 230V models

o

15-Amp dedicated single phase circuit. Additional

appliances on the same circuit may cause the current to
exceed the rating of the circuit breaker or fuse.

o

Frequency must agree with the serial tag rating.

Improper frequency will damage the compressor.



Step 2. Rotary Vane Vacuum Pump Requirements

• 1 Liter, 2.5 Liter, 4.5 Liter models

o

88 liters per minute minimum displacement

o

.002 mBar ultimate vacuum

• 6 Liter, 12 Liter, 18 Liter models

o

98 liters per minute minimum displacement

(144 liters per minute recommended)

o

.002 mBar ultimate vacuum

• Plug Requirements on vacuum pump power cord

o

230V models- reverse IEC plug

• Hose barb connection to fit .75-inch inside diameter hose

Reverse IEC plug

UK plug

EU plug