Labconco FreeZone User Manual

Page 4

Advertising
background image

Page 3 of 5

Pull

• Fill the pump with oil so the level is between the lines on the sight gauge on the end of

the pump housing.

• Verify that the vacuum pump voltage setting agrees with the Freeze Dryer voltage

shown on the serial number label. Reconfigure if necessary. See pump instructions.

• Turn the vacuum pump power switch ON.


• Set the gas

ballast to
closed or
minimum
position.





Step 4. Vacuum Pump Connections

• Benchtop

models

o

Position the vacuum pump within 2 feet

of the .75-inch diameter port on the
back of the cabinet.

o

Attach one end of the vacuum hose to

the port on the Freeze Dryer and the
other end to the inlet port of the
vacuum pump. Clamp both ends
securely.

o

Connect the vacuum pump power cord

into the receptacle on the back of the
Freeze Dryer.

• Console

models

o

CAUTION - Make sure that power to

the Freeze Dryer is disconnected.

o

Remove the front lower panel from the

Freeze Dryer by inserting a screw
driver into the slot below the center of
the front panel. Pivot the screwdriver
handle downward while pulling the
bottom of the panel outward.

Push down