Remove used oil – Labconco PrimeMate Oil Change System 73570 Series User Manual

Page 17

Advertising
background image

Chapter 4: Using Your PrimeMate Oil Change System

Product Service 1-800-522-7658

13

replenished. It is important to monitor the oil level in the vacuum
pump sight glass. Turn the fill switch off when the oil level is
satisfactory.

3. Drain Light

This light will illuminate green when the drain pump is operating.

4. Waste Full Light

This light illuminates red to indicate that the dirty oil container inside
the PrimeMate Oil Change System is full. The drain pump and green
drain light both turn off when this container is full.

5. Fill Light

This light illuminates green when the fill pump is in operation.

Remove Used Oil


Obtain access to the used oil container by removing the cover. Grasp the handle
and pull straight upward. The used oil container is on the left side. Unscrew the
cap and remove the cap/float switch assembly. Remove the container.

IMPORTANT NOTE: THE DISPOSAL OF SUBSTANCES
USED IN CONNECTION WITH THIS EQUIPMENT MAY BE
GOVERNED BY VARIOUS FEDERAL, STATE OR LOCAL
REGULATIONS. ALL USERS OF THIS EQUIPMENT ARE
URGED TO BECOME FAMILIAR WITH ANY
REGULATIONS THAT APPLY IN THE USER’S AREA
CONCERNING THE DUMPING OF WASTE MATERIALS IN
OR UPON WATER, LAND OR AIR AND TO COMPLY
WITH SUCH REGULATION.

Oil may be removed from the container and the container re-used, or if new
oil is purchased from Labconco and the new oil container is used in the
PrimeMate Oil Change System, the empty new oil container may be moved
to the used oil container position.

Return the Vacuum Pump to Service
After an Oil Change


Turn off the power switch on the PrimeMate Oil Change System. Disconnect
the fill hose from the vacuum pump by pressing the tab on the side of the
coupling body. The hose from the PrimeMate Oil Change System must be
disconnected from the vacuum pump when the oil change is completed. This
will prevent the loss of oil and subsequent damage to the vacuum pump in
case the hose is damaged or develops a leak.

)

Advertising