Installation of Gloves to the Glove Ports

With the thumbs up and in a right/left orientation, secure the gloves in place
on the glove ports by stretching the beaded glove cuff into the glove port
groove nearest the window. Install the separate 8" diameter O-ring over the
gloves, into the outer groove of the glove port surface. Stainless steel band
clamps are provided for securing the separate O-ring into the glove port
groove. Replacement gloves and parts are listed in Chapter 7.

Figure 3-4

Glove Installation

Connecting the Electrical Supply Source to the
Glove Box

115V, 60 Hz; 100V, 50 Hz; or 100V, 60 Hz Models

Connect the power cord supplied to the IEC electrical supply plug located on
the back of the glove box. 115V or 100V models are rated at 8 amps total.
The maximum circuit load for the interior electrical duplex is 5 amps.

230V, 50 Hz or 230V, 60 Hz Models

The same procedure applies for the 230V international models except the
cords are shipped without an attached plug. Install the appropriate plug for
your electrical specifications per local codes. The 230V, 50 Hz or 230V, 60
Hz models are both rated at 4 amps total. The interior electrical duplex is
rated at 2 amps.