Chapter 6: Maintaining Your Glove Box

Product Service 1-800-522-7658

34

7. Install the new pressure relief valve with the gasket to the inside.

Tighten the large nut.

8. Turn the glove box ON and re-certify the glove box.

9. Refer to Appendix A, Appendix B and Chapters 3 and 7 for

installation diagrams.

Service Valve and Pressure Gauge Replacement

1. Decontaminate the glove box interior, if required.

2. Wear appropriate protective equipment as determined by your Safety

Officer.

3. Turn the glove box and vacuum pump OFF.

4. Remove the valve or pressure gauge by loosening the large nut for the

valve or hardware for the pressure gauge. Remove the pressure hose
connected to the pressure gauge.

5. Replace with the new valve with the gasket inside or replace with the

new pressure gauge.

6. Tighten the large nut for the valve or the hardware for the pressure

gauge and re-attach the pressure hose to the pressure gauge.

7. Turn the glove box ON and re-certify.

8. Refer to Appendix A and Chapters 3 and 4 for installation diagrams.