Pressure relief valve test procedure, Other equipment – Labconco Precise Controlled Atmosphere and Basic Glove Boxes 5220131 User Manual

Page 58

Advertising
background image

Appendix G: Helium Leak Test Procedure

Product Service 1-800-522-7658

54

Pressure Relief Valve Test Procedure

1. See that pressure relief valve opens when pressures and vacuum

exceed +/- 6" to 10" w.g. Record data. Check that valve reseals
within 2" w.g. of opening pressure and opening vacuum. Pressure
relief valve is standard on Precise Controlled Atmosphere Glove
Boxes.

Other Equipment

1. For a less critical helium leak measurement using a lower cost leak

detector, Labconco recommends using Restek (www.restek.com) or
other instrumentation available from a laboratory supply dealer.