ROUTINE MAINTENANCE

15

Gloves

At least every three months (more often if subjected to heavy use), inspect the gloves for

signs of damage or wear. For replacement, refer to page 11.

Transfer Chamber Inner and Outer Door

Periodically check the closing tension of the door latches. Closing the latch handle should

require some firm force, indicating compression of the door seals. Seals, compression, and
latching force can be adjusted as follows:

1. Close and latch the door.
2. Using a 1/2" open-end wrench, turn the door adjustment screw (located between the door

and the latch bar) clockwise to increase seal compression and latching force. Turn screw
counterclockwise to decrease compression. Refer to figure 5.
























Figure 5